HP8648 emulation mode
1GP71_0E
134
Rohde & Schwarz
Status
Command / Function / Remark / Limitation
*SAV value
Save an instrument setup.
*SRE value
Set / Query the bits of the status enable register.
*STB?
Query the bits of the status register.
*TST?
Execute the instruments self test
*WAI
Wait until al commands are completed.
Table 129: Common commands
•
AM Subsystem
Status
Command / Function / Remark / Limitation
[SOURce:]AM[:DEPTh] value
Set /Query the modulation depth.
Value:
0 … 100 PCT
Default:
30 PCT
[SOURce:]AM:STATe value
Set / Query the (activation) state of the modulation.
Value:
ON | OFF | 1 | 0
Default:
OFF
[SOURce:]AM:SOURce value
Select / Query the signal source of the modulation.
Value:
INTernal[1], EXTernal | EXTernal, INTernal[1] | INTernal2
Default:
INTernal
[SOURce:]AM:EXTernal:COUPling value
Select / Query the coupling type, when modulation source is set to external.
Value:
AC | DC
Default: