HP8340 emulation mode
1GP71_0E
70
Rohde & Schwarz
RB[value]
Enable KNOB in remote-control interface, optionally modify active parameter by
<value>.
SN value
Select stepped / manual sweep mode, optionally sets number of sweep steps from
start to stop to <value>.
IF
Increment frequency.
RS
Reset sweep.
TS
Take sweep.
TL value
Set time limit to <value>
Remarks:
SM* allows only time limit values, that are multiples of 10ms.
Sweep time is computed as
tsweep = dwell * count
dwell
is set to 10ms,
count
is computed.
As time limit is the minimal sweep time, time limit is computed as
tlimit = dwell * count(min)
AS0
Select foreground (backup) register.
AS1
Select background (backup) register.
FP
Select fast phaselock.
AT
Control step attenuator independently.
TI value
Test GPIB interface.
Value:
0x00 … 0xFF
Default:
0
Remark:
Response depends on firmware release of HP8340.
SM* always responds <value>.
Table 60: GPIB commands
•
Others / Unknown
Status
Command / Function / Remark / Limitation
SHPM