RC3102 emulation mode
1GP71_0E
167
Rohde & Schwarz
FS value
Set frequency step size in fixed and sweep mode.
Value:
value in HZ | KZ | MZ | GZ
Default:
12.5 KZ
FU
Increment fixed frequency by value specifed with FS command.
FD
Decrement fixed frequency by value specifed with FS command.
Table 161: Frequency commands
•
Level
Status
Command / Function / Remark / Limitation
AP value
Select amplitude function, optionally set amplitude.
Value:
value in DB | VO | MV | UV | NV
Default:
-30.0 DB
AR value
Set amplitude offset.
Value:
value in DB | VO | MV | UV | NV
Default:
0.0 DB
AS value
Set amplitude step size.
Value:
value in DB | VO | MV | UV | NV
Default:
3.0 DB
AU
Increment amplitude by value specifed with AS command.
AD
Decrement amplitude by value specifed with AS command.
OP1
Turn Rf output on.
OP0
Turn Rf ouput off.
Table 162: Level commands