HP8643 emulation mode
1GP71_0E
112
Rohde & Schwarz
Status
Command / Function / Remark / Limitation
*OPT?
Query the installed options.
*RCL value
Recall an instrument setup.
Value:
0 … 9
*RST
Reset the instrument to a default state.
*SAV value
Save an instrument setup.
Value:
0 … 9
*SRE value
Set / Query the bits of the status enable register.
*STB?
Query the bits of the status register.
*TST?
Execute the instruments self test
*WAI
Wait until al commands are completed.
Table 105: Common commands
•
AM Subsystem
Status
Command / Function / Remark / Limitation
[SOURce:]AM[:DEPTh] value
Set / Query the rf carrier modulation depth.
Value:
value [PCT | %] | UP | DOWN | MINimum | MAXimum
Default:
0.0 PCT
Remark:
% is not supportted.
[SOURce:]AM[:DEPTh]:STEP[:INCRement] value
Set / Query the rf carrier modulation depth step size.
Value:
value [PCT | %] | MINimum | MAXimum
Default:
1.0 PCT
Remark:
% is not supportted.