HP8662 emulation mode
1GP71_0E
150
Rohde & Schwarz
13
HP8662 emulation mode
General
In this language emulation mode, the emulation software in the SM* emulates
the Pre-SCPI language, implemented in Agilent’s signal generators of type
o
HP8662A
o
HP8663A
The emulation software is supported on signal generators of type
o
SMA100
The emulation software simulates the HP8662 data model in the execution unit
and transforms the HP8662 parameters into the SM* parameters.
Figure 10: HP8662 emulation model
This mapping occurs only in one direction: Setting commands that arrive via
the remote-control interface modify the HP8662 parameters. These
parameters are converted into SM* parameters and written to the database.
Modifiying the SM* parameters in any other manner, e.g., via manual operation
at the frontpanel, causes no changes to the HP8662 parameters of the
emulation software.
While this is insignificant in pure remote-control operation, it can lead to
inconsistencies and undesirable effects or malfunctions when you use
both manual and remote control in this language emulation mode.
GPIB functions
The following table shows the GPIB functions, supported by the HP8662.
Status
Command
Function / Remark / Limitation