HP8373 emulation mode
1GP71_0E
102
Rohde & Schwarz
Status
Command / Function / Remark / Limitation
[SOURce[1]:]AM[:DEPTh] value
Set / Query the modulation depth.
Value:
0.0 … 100.0 [PCT | %] | UP | DOWN | MINimum | MAXimum | DEFault
Default:
6 PCT
Remark:
% is not supported
UP, DOWN, MINimum, MAXimum, DEFault are not supported.
[SOURce[1]:]AM[:DEPTh]:STEP[:INCRement] value
[SOURce[1]:]AM:INTernal:FREQuency value
Set / Query the modulation rate.
Value:
0.5 Hz … 100.0 KHz in [HZ | KHZ] | UP | DOWN | MINimum | MAXimum | DEFault
Default:
5.0 KHz
Remark:
LF Generator 1 is used
UP, DOWN, MINimum, MAXimum, DEFault are not supported.
[SOURce[1]:]AM:INTernal:FREQuency:STEP[:INCRement] value
[SOURce[1]:]AM:INTernal:FUNCtion value
[SOURce[1]:]AM:SENSitivity value
[SOURce[1]:]AM:SOURce value
Select / Query the signal source of the modulation.
Value:
INTernal | FEED | EXTernal
Default:
EXTernal
Remark:
INTernal, FEED, EXTernal are not supported
LF1 corresponds to INTernal
EXT1 corresponds to EXTernal
[SOURce[1]:]AM:STATe value
Set / Query the (activation) state of the modulation.
Value:
ON | OFF | 1 | 0
Default:
OFF
[SOURce[1]:]AM:TYPE value
[SOURce[1]:]FM:COUPling value