HP8340 emulation mode
1GP71_0E
59
Rohde & Schwarz
Select start/stop frequency sweep mode, optionally set start frequency.
Value:
freqmin … freqmax HZ | KZ | MZ | GZ
Default:
freqmin
FB[value]
Select start/stop frequency sweep mode, optionally set stop frequency.
Value:
freqmin … freqmax HZ | KZ | MZ | GZ
Default:
freqmax
CF[value]
Select center/delta frequency sweep mode, optionally set center frequency.
Value:
freqmin … freqmax HZ | KZ | MZ | GZ
Default:
0.5 * (f freqmax)
DF[value]
Select center/delta frequency sweep mode, optionally set delta frequency.
Value:
0 … freqmax HZ | KZ | MZ | GZ
Default:
freqmax - freqmin
SHST[value]
Zoom in/out in center/deltay frequency sweep mode.
Value:
freqmin … freqmax HZ | KZ | MZ | GZ
Default:
0.5 * (f freqmax)
Remark:
Acts like CF[value]
Frontpanel knob and step keys are not linked to center frequency and delta
frequency.
Commands RB[value] and UP/DN are not linked to center frequency and delta
frequency.
CW[value]
Select fixed frequency mode, optionally set fixed frequency.
Value:
freqmin … freqmax HZ | KZ | MZ | GZ
Default:
0.5 * (f freqmax)
SHCW[value]
Select fixed frequency mode, optionally set fixed frequency.
Value: