Trying Out the Instrument
R&S
®
RTO2000
77
Getting Started 1332.9719.02 ─ 10
6.13
Saving Data
After a measurement with the R&S
RTO, you can save the resulting waveform
data for further evaluation or comparison. You can also save measurement
results, and device settings to repeat or restore previous measurements.
●
●
"Saving data of an acquisition series"
●
●
"Saving and restoring device settings"
Saving waveform data
1. Press the [Save Recall] key on the front panel (in the Setup area on the left).
2. Select "Waveform".
3. Check the "Source".
4. Set "Scope" to "Full Waveform".
Tip:
If a cursor, zoom or measurement gate is defined, you can use these set-
tings to export only a part of the waveform.
5. Under "Save to file", tap "Save As".
6. The file selection dialog box shows the default storage directory:
C:\Users\Public\Public Documents\Rohde-Schwarz\RTx\
RefWaveforms
7. Tap the keyboard icon on the right of the "File Name" field.
8. Enter
Waveform1
on the online keyboard.
9. Tap "ENTER" to close the online keyboard.
10. Select the file type: "*.bin".
11. Tap "Save".
The waveform data is saved to the files
Waveform1.Wfm.bin
and
Waveform1.bin
in the selected directory.
Saving Data