Trying Out the Instrument
R&S
®
RTO2000
41
Getting Started 1332.9719.02 ─ 10
8. Tap the "Undo" icon on the toolbar repeatedly until the trigger mode is "Auto"
and the trigger source is "CH1".
9. Press the [Slope] key to toggle the trigger slope.
Watch the waveform and the "Trigger" settings.
6.2
Acquiring Data
Although you are using only one input channel, you can acquire the calibration
data using three different arithmetic methods and display the results in separate
waveforms.
1. In the [Horizontal] area, press the [Acquisition] key.
2. In the "Acquisition" tab of the "Horizontal" dialog box, enable all three wave-
forms ("Wfm1", "Wfm2", "Wfm3").
If you cannot see these settings, enable "Multi waveform" in the lower right
corner of the dialog box.
3. For Wfm1, select the "Mode" =
Sample
and "Wfm Arithmetic" = "Off".
Acquiring Data