Instrument Tour
R&S
®
RTO2000
31
Getting Started 1332.9719.02 ─ 10
You can select if the knob changes the position or the reference point in "File"
menu > "Frontpanel Setup" > "Knobs". To set the value to zero, press the knob.
The current value is shown in the input box in the lower right corner of the screen.
"Horizontal position" defines the time distance of the reference point from the zero
point of the diagram. Turn clockwise to move the waveform to the right.
"Reference point" defines the position of the reference point on the screen. Turn
clockwise to move it to the right. The reference point marks the rescaling center
of the time scale. It is indicated by a gray triangle outline at the top of the dia-
gram. If you modify the time scale, the reference point remains fixed on the
screen, and the scale is stretched or compressed to both sides of the reference
point.
[Scale]
The rotary knob adjusts the time scale for all signals. The time scale is also
known as timebase.
Turn clockwise to stretch the waveforms. Doing so, the scale value
time/div
decreases.
Press the knob to toggle between coarse and fine scale adjustment.
5.3.4
Vertical Controls
The keys and knobs in the Vertical functional block select a signal and adjust the
vertical scale and position of the selected signal.
Keys and Controls