A math waveform is configured that uses the "Mag(FFT(x))" operator with Ch1
as source. The spectrum waveform is displayed in a new diagram.
3. To measure the spectrum on the math channel, tap the "Measurement" icon
on the toolbar.
4. Tap the spectrum waveform.
The channel power is displayed below the FFT diagram.
5. Double-tap the spectrum waveform.
The sidebar opens, showing main FFT settings.
6. Set the "Center frequency" to
1
MHz
.
The instrument adjusts the frequency span automatically.
Close the sidebar.