![R&S RTO2000 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/rands/rto2000/rto2000_getting-started_3208181025.webp)
Instrument Tour
R&S
®
RTO2000
25
Getting Started 1332.9719.02 ─ 10
If grounding is
not
ensured by the mains system, ground the oscilloscope using
the protective earth conductor on the front panel and an appropriate cable.
The AC main power switch also interrupts the power supply of the OCXO (option
OCXO Reference Frequency, R&S RTO-B4).
When you power up the instrument, be sure to comply with the warm-up phase
specified in the data sheet before you start measurements.
USB
Two USB type A connectors that comply with standard USB 3.0. They are used to
connect devices like keyboard, mouse, printer, and flash drive to store and reload
instrument settings and measurement data.
Note:
Electromagnetic interference (EMI) can affect the measurement results. To
avoid any impact, do not use USB connecting cables exceeding 1
m.
LAN
8-pin RJ-45 connector used to connect the instrument to a Local Area Network
(LAN). It supports up to 1000 Mbit/s (10/100/1000BASE-T Ethernet).
USB Device
USB 3.0 interface of type B (device USB), to be used for remote control of the
instrument.
DisplayPort
DisplayPort connector for an external monitor or projector. It supports DisplayPort
version 1.1a.
DVI-D
Digital connector for an external monitor or projector. The monitor shows the com-
plete content of the instrument's screen.
Ext Trigger In
The BNC connector for external trigger input is used to control the measurement
by an external signal. The input impedance can be selected in the trigger configu-
ration, the values are 50
Ω
and 1
M
Ω
. The trigger level can be set from -5
V to
5
V. The maximum input voltage is 30
V
RMS at 1
M
Ω
input impedance and
7
V
RMS at 50
Ω
input impedance.
Rear Panel