Trying Out the Instrument
R&S
®
RTO2000
70
Getting Started 1332.9719.02 ─ 10
4. In the "Window" tab below, define the search conditions:
a) In the "Upper level" field, enter the result of the "High" level measurement.
b) In the "Lower level" field, enter the result of the "Low" level measurement.
c) As "Vertical condition", select "Exit" to find values that are outside the
range defined by the high and low levels.
5. Select "Enable" to start the continuous search on the acquired data.
6. Close the "Search" dialog box.
7. In the "Search Results" box, select "Show search zoom windows".
The acquisition stops, and the detected overshoots of the last acquisition are
listed in the search result table. The search zoom window shows the last
result that was found. Vertical lines indicate the time values for which a result
was found.
Performing a Search