Remote Commands
R&S
®
FSMR3000
430
User Manual 1179.0116.02 ─ 02
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor
<Factor>
Parameters:
<Factor>
1 to 100 PCT
*RST:
10
Default unit: PCT
Example:
//Couple frequency step size to span and define a step size fac-
tor
FREQ:CENT:STEP:LINK SPAN
FREQ:CENT:STEP:LINK:FACT 20PCT
[SENSe:]FREQuency:CW:AFC
<state>
Executes an automatic signal search and activates a function to track a drifting signal.
If the measured signal frequency is too far away from the tuned center frequency and
the level is above the threshold, the center frequency of the R&S
FSMR3 will be
changed to the new signal frequency.
Parameters:
<State>
Example:
FREQ:CW:AFC ONCE
12.5.4
Configuring the Vertical Axis (Amplitude, Scaling)
The following commands are required to configure the amplitude and vertical axis set-
tings in a remote environment.
................................................................................................ 430
...................................................................................431
.....................................................................................433
12.5.4.1
Amplitude Settings
Remote commands exclusive to amplitude configuration:
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
........................................................... 430
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
........................................................431
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
........................................... 431
CALCulate<n>:MARKer<m>:FUNCtion:REFerence
This command matches the reference level to the power level of a marker.
If you use the command in combination with a delta marker, that delta marker is turned
into a normal marker.
Configuring the Measurement