General Instrument Setup
R&S
®
FSMR3000
304
User Manual 1179.0116.02 ─ 02
........................................................................................... 304
..........................................................................................305
NONE
Uses the current RF signal at the input, i.e. no calibration signal (default).
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
Calibration Frequency RF
Uses the internal calibration signal as the RF input signal.
Remote command:
DIAGnostic:SERVice:INPut[:SELect]
DIAGnostic:SERVice:INPut:PULSed:CFRequency
Spectrum ← Calibration Frequency RF
Defines whether a broadband or narrowband calibration signal is sent to the RF input.
"Narrowband"
Used to calibrate the absolute level of the frontend at 64 MHz.
"Broadband"
Used to calibrate the IF filter.
Remote command:
DIAGnostic:SERVice:INPut:RF[:SPECtrum]
Frequency ← Calibration Frequency RF
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration (max. 64
MHz).
For narrowband signals, 64
MHz is sent.