Common Analysis and Display Functions
R&S
®
FSMR3000
236
User Manual 1179.0116.02 ─ 02
5. In the file selection dialog box, select the storage location and file name for the
export file.
6. If necessary, change the decimal separator to be used for the ASCII export file.
7. Select the data format of the ASCII file.
8. Select "Save" to close the dialog box and export the data to the file.
8.5.3
Reference: ASCII File Export Format
Trace data can be exported to a file in ASCII format for further evaluation in other appli-
cations. This reference describes in detail the format of the export files for result data.
The file consists of the header information (general configuration of the measurement)
and the measurement results. Optionally, the header can be excluded from the file.
The data of the file header consist of three columns, each separated by a semicolon:
parameter name; numeric value; basic unit. The data section starts with the keyword
"Trace <n>" (<n> = number of stored trace), followed by the measured data in one or
several columns (depending on the measurement) which are also separated by a sem-
icolon.
The results are output in the same order as they are displayed on the screen: window
by window, trace by trace, and table row by table row.
Generally, the format of this ASCII file can be processed by spreadsheet calculation
programs, e.g. MS Excel. Different language versions of evaluation programs may
require a different handling of the decimal point. Thus you can define the decimal sep-
arator to be used (decimal point or comma).