Remote Commands
R&S
®
FSMR3000
501
User Manual 1179.0116.02 ─ 02
*RST:
SINGle
Manual operation:
See
"Export all Traces and all Table Results"
FORMat:DEXPort:XDIStrib
<XDistribution>
Defines how the x-values of the trace are determined in the frequency domain.
Parameters:
<XDistribution>
STARtstop | BINCentered
BINCentered
The full measurement span is divided by the number of mea-
surement points to obtain
bin
s. The x-value of the measurement
point is defined as the x-value at the center of the bin (bin/2).
STARtstop
(Default): The x-value of the first measurement point corre-
sponds to the starting point of the full measurement span. The x-
value of the last measurement point corresponds to the end
point of the full measurement span. All other measurement
points are divided evenly between the first and last points.
Example:
FORM:DEXP:XDIS BINC
FORMat:DIMPort:TRACes
<Selection>
This command selects the data to be included in a data import file (see
Parameters:
<Selection>
SINGle | ALL
SINGle
Only a single trace is selected for import, namely the one speci-
fied by the
mand.
ALL
Imports several traces at once, overwriting the existing trace
data for any active trace in the result display with the same trace
number. Data from the import file for currently not active traces
is not imported.
The <trace> parameter for the
on page 501 command is ignored.
*RST:
SINGle
MMEMory:LOAD<n>:TRACe
<Trace>, <FileName>
This command imports trace data from the specified window to an ASCII file.
Suffix:
<n>
.