Remote Commands
R&S
®
FSMR3000
506
User Manual 1179.0116.02 ─ 02
Command syntax
Evaluation method
ACV[:TDOMain]
AC-Video time domain
ACV:AFSpectrum
AC-Video spectrum
AM[:ABSolute][:TDOMain]
RF time domain
AM:RELative[:TDOMain]
AM time domain
AM:RELative:AFSPectrum
AM spectrum (relative)
FM[:TDOMain]
FM time domain
FM:AFSPectrum
FM spectrum
PM[:TDOMain]
PM time domain
PM:AFSPectrum
PM spectrum
SPECtrum
RF spectrum
Suffix:
<n>
.
irrelevant
Parameters:
<TraceMode>
WRITe
Overwrite mode: the trace is overwritten by each sweep. This is
the default setting.
AVERage
The average is formed over several sweeps.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
FSMR3 saves the sweep result in the trace
memory only if the new value is greater than the previous one.
MINHold
The minimum value is determined from several measurements
and displayed. The R&S
FSMR3 saves the sweep result in the
trace memory only if the new value is lower than the previous
one.
VIEW
The current contents of the trace memory are frozen and dis-
played.
OFF
Hides the selected trace.
*RST:
WRITe,OFF,OFF,OFF,OFF,OFF
Example:
ADEM:AM AVER,MAXH,MINH,OFF,OFF,OFF
Determines average, max hold and min hold values simultane-
ously for the traces 1-3 of the RF time domain evaluation.
ADEM:AM WRIT,OFF,OFF,OFF,OFF,OFF
Determines only the current measurement values for trace 1.
ADEM:AM OFF,OFF,OFF,OFF,OFF,OFF
Switches AM demodulation off.