Remote Commands
R&S
®
FSMR3000
470
User Manual 1179.0116.02 ─ 02
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
<Threshold>
When the reference level is adjusted automatically using the
on page 471 command, the internal attenuators and the preamplifier are also
adjusted. In order to avoid frequent adaptation due to small changes in the input signal,
you can define a hysteresis. This setting defines a lower threshold the signal must fall
below (compared to the last measurement) before the reference level is adapted auto-
matically.
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:LOW 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18
dBm.
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
<Threshold>
When the reference level is adjusted automatically using the
on page 471 command, the internal attenuators and the preamplifier are also
adjusted. In order to avoid frequent adaptation due to small changes in the input signal,
you can define a hysteresis. This setting defines an upper threshold the signal must
exceed (compared to the last measurement) before the reference level is adapted
automatically.
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:UPP 2
Example:
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22
dBm.
[SENSe:]ADJust:CONFigure:SMODe
<Search Mode>
Determines the search mode for the automatic measurement performed to determine
the optimal measurement configuration.
Parameters:
<Search Mode>
FAST | POPTimzed
FAST
The measurement is optimized for speed.
POPTimzed
The measurement is optimized to analyze pulse signals ade-
quately.
Example:
ADJ:CONF:SMOD POPT