Remote Commands for Synchronizing Parameters
..............................................626
12.8.1
Configuring the Reference Frequency
.........................................................................................593
............................................................................................. 593
............................................................................................. 593
[SENSe:]ROSCillator:EXTernal:FALLback
........................................................................ 594
SOURce<si>:EXTernal<ext>:ROSCillator:EXTernal:FREQuency
.........................................594
.......................................................................................... 595
........................................................................................ 595
[SENSe:]ROSCillator:SOURce:EAUTo?
........................................................................... 596
.........................................................................................596
[SENSe:]ROSCillator:LBWidth
<Bandwidth>
Defines the loop bandwidth, that is, the speed of internal synchronization with the refer-
ence frequency. The setting requires a compromise between performance and increas-
ing phase noise.
For a variable external reference frequency with a narrow tuning range (+/- 0.5 ppm),
the loop bandwidth is fixed to 0.1 Hz and cannot be changed.
Parameters:
<Bandwidth>
0.1 Hz | 1 Hz | 3 Hz | 10 Hz | 30 Hz | 100 Hz | 300 Hz
The possible values depend on the reference source and tuning
range (see
Default unit: Hz
Example:
ROSC:LBW 3
Manual operation:
See
[SENSe:]ROSCillator:O100
<State>
[SENSe:]ROSCillator:O640
<State>
This command turns the output of a reference signal on the corresponding connector
("Ref Output") on and off.
[SENSe:]ROSCillator:O100
: Provides a 100
MHz reference signal on correspond-
ing connector.
[SENSe:]ROSCillator:O640
: Provides a 640
MHz reference signal on correspond-
ing connector.
Parameters:
<State>
ON | OFF | 1 | 0
Configuring the R&S
FSMR3