Remote Commands
R&S
®
FSMR3000
437
User Manual 1179.0116.02 ─ 02
OFF | 0
Switches FM demodulation mode off
ON | 1
Switches FM demodulation mode on
*RST:
0
Example:
ADEM ON
[SENSe:]ADEMod:MTIMe
<Time>
This command defines the measurement time for Analog Modulation Analysis.
Parameters:
<Time>
*RST:
62.5us
Default unit: S
Example:
ADEM:MTIM 62.5us
Sets the measurement time to 62.5 µs.
Manual operation:
See
[SENSe:]ADEMod:RLENgth
[SENSe:]ADEMod:SET
<SampleRate>, <RecordLength>, <TriggerSource>,
<TriggerSlope>, <OffsetSamples>, <NoOfMeas>
This command configures the analog demodulator of the instrument.
Parameters:
<SampleRate>
numeric value
The frequency at which measurement values are taken from the
A/D-converter and stored in I/Q memory.
*RST:
8 MHz
Default unit: HZ
<RecordLength>
Number of samples to be stored in I/Q memory.
Range:
1 to 400001 with AF filter or AF trigger active, 1 to
480001 with both AF filter and AF trigger deactive
*RST:
501)
<TriggerSource>
IMMediate | EXTernal | EXT2 | EXT3 | IFPower | RFPower| AF
| AM | AMRelative | FM | PM
Note:
After selecting IF Power, the trigger threshold can be set
with the
TRIGger[:SEQuence]:LEVel:IFPower
*RST:
IMMediate
Configuring the Measurement