Remote Commands
R&S
®
FSMR3000
469
User Manual 1179.0116.02 ─ 02
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
............................................................... 470
[SENSe:]ADJust:CONFigure:SMODe
...............................................................................470
[SENSe:]ADJust:CONFigure:TRIGger
..............................................................................471
..........................................................................................471
[SENSe:]ADJust:SCALe[:Y]:AUTO[:CONTinuous]
............................................................. 471
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Center frequency
●
Reference level
Example:
ADJ:ALL
[SENSe:]ADJust:CONFigure:LEVel:DURation
<Duration>
In order to determine the ideal reference level, the R&S
FSMR3 performs a measure-
ment on the current input data. This command defines the length of the measurement
if
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
<Mode>
In order to determine the ideal reference level, the R&S
FSMR3 performs a measure-
ment on the current input data. This command selects the way the R&S
FSMR3 deter-
mines the length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
FSMR3 determines the measurement length automati-
cally according to the current input data.
MANual
The R&S
FSMR3 uses the measurement length defined by
[SENSe:]ADJust:CONFigure:LEVel:DURation
on page 469.
*RST:
AUTO
Configuring the Measurement