Remote Commands
R&S
®
FSMR3000
471
User Manual 1179.0116.02 ─ 02
[SENSe:]ADJust:CONFigure:TRIGger
<State>
Defines the behavior of the measurement when adjusting a setting automatically (using
SENS:ADJ:LEV ON
, for example).
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
[SENSe:]ADJust:FREQuency
This command sets the center frequency to the frequency with the highest signal level
in the current frequency range.
Example:
ADJ:FREQ
[SENSe:]ADJust:LEVel
This command initiates a single (internal) measurement that evaluates and sets the
ideal reference level for the current input data and measurement settings. This ensures
that the settings of the RF attenuation and the reference level are optimally adjusted to
the signal level without overloading the R&S
FSMR3 or limiting the dynamic range by
an S/N ratio that is too small.
Example:
ADJ:LEV
[SENSe:]ADJust:SCALe[:Y]:AUTO[:CONTinuous]
<State>
Activates automatic scaling of the y-axis in all diagrams according to the current mea-
surement results. Currently auto-scaling is only available for AF measurements. RF
power and RF spectrum measurements are not affected by the auto-scaling.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
SENS:ADJ:SCAL:Y:AUTO ON
Manual operation:
See
12.5.10
Capturing Data and Performing Sweeps
The following commands are required to capture data.
Configuring the Measurement