Remote Commands
R&S
®
FSMR3000
524
User Manual 1179.0116.02 ─ 02
CALCulate<n>:MARKer<m>:FUNCtion:MDEPth[:STATe]
<State>
This command turns the AM Modulation Depth measurement on and off.
To work correctly, the measurement requires an AM modulated signal.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
CALCulate<n>:MARKer<m>:FUNCtion:ADEMod:RFFRequency[:RESult<t>]?
Queries the result of the RF frequency counter measurement.
Suffix:
<n>
.
<m>
<t>
Return values:
<Value>
<numeric value>
Example:
CALC:MARK:FUNC:ADEM:RFFR?
Usage:
Query only
CALCulate:MARKer:FUNCtion:ADEMod:AM[:RESult<t>]?
<MeasType>
CALCulate:MARKer:FUNCtion:ADEMod:FM[:RESult<t>]?
<MeasType>
CALCulate<n>:MARKer<m>:FUNCtion:ADEMod:PM[:RESult<t>]?
<MeasType>
This command queries the current value of the demodulated signal for the specified
trace (as displayed in the Result Summary in manual operation).
Note that all windows with the same evaluation method have the same traces, thus the
window is irrelevant.
Suffix:
<n>
.
<m>
irrelevant
<t>
Query parameters:
<MeasType>
PPEak | MPEak | MIDDle | RMS
PPEak
Postive peak (+PK)
MPEak | NPEak
Negative peak (-PK)
Analyzing Measurements (Basics)