Remote Commands
R&S
®
FSMR3000
431
User Manual 1179.0116.02 ─ 02
Suffix:
<n>
.
<m>
Example:
CALC:MARK2:FUNC:REF
Sets the reference level to the level of marker 2.
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
<ReferenceLevel>
This command defines the reference level (for all traces in all windows).
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
0 dBm
Default unit: DBM
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<Offset>
Range:
-200 dB to 200 dB
*RST:
0dB
Default unit: DB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation:
See
"Shifting the Display (Offset)"
12.5.4.2
Configuring the Attenuation
......................................................................................... 432
INPut<ip>:ATTenuation
<Attenuation>
This command defines the total attenuation for RF input.
Configuring the Measurement