R&S FSH
Cable Measurements
1145.5973.12 4.121
E-14
Press the FREQ key.
Using either the numeric keys, the cursor keys or
the rotary knob, set the desired frequency.
Confirm the entry with the ENTER key or the
CENTER FREQ softkey.
Calibrating the test setup
The test setup must be calibrated before any measurements are performed. To perform calibration, a
SHORT is required at the output of the 1 m measurement cable. An OPEN can be used instead of a
SHORT. However, if an OPEN is used, greater measurement uncertainties must be expected as an
OPEN is not defined as precisely as a SHORT.
Note:
The reference plane must be the output of the 1 m measurement cable; i.e. the
measurement cable may not be dispensed with. If the output of the VSWR bridge is used
as the reference plane, the DTF results are useless.
Press the DTF CAL softkey.
The R&S FSH opens a text window which prompts the
user to terminate the measurement cable with a
SHORT.
Firmly screw the SHORT to the output end of the
measurement cable.
Press the CONTINUE softkey to start the SHORT
calibration.
Calibration can be aborted by pressing CANCEL.
While SHORT calibration is in progress, the R&S FSH
outputs the message "Calibrating SHORT, please
wait...".
Calibration can be aborted with the CANCEL softkey.