Two-port measurements with the tracking generator
R&S FSH
1145.5973.12 4.112
E-14
To shift in the y direction:
Press the MOVE Y softkey.
Set a value from -50% to +50% by using the rotary
knob or cursor keys or by entering a numeric value.
Confirm the selection with the ENTER key or the
MOVE Y softkey.
To enlarge an area:
Press the ZOOM IN softkey.
The selected window area will be enlarged by the
zoom factor that has been set. You can fine-adjust the
zoom window by using the MOVE X and MOVE Y as
described.
To deactivate enlargement:
Press the ZOOM OUT key.
Measuring the group delay when measuring reflection
For more information, see section "Measuring the group delay when measuring transmission".
Press the MEAS MODE softkey.
Using the rotary knob or the cursor keys, select
GROUP DELAY from the menu and confirm by
pressing either the MEAS MODE softkey or the
ENTER key.
A field for entering the aperture will open and indicate
the aperture value currently selected. The default
setting is an aperture width of ten
test points.
Acceptable entry values are whole numbers from 1 to
300. For example, an entry of APERTURE
= 5 tells the
system to use the phase values of test points n-3 and
n+2 when performing the calculation at test point n.