Two-port measurements with the tracking generator
R&S FSH
1145.5973.12 4.86
E-14
Press the FREQ key.
Using the number keys, enter the center frequency (380 MHz in this example).
Press the SPAN key.
Using the number keys, enter the span (25 MHz in this example).
Scalar transmission measurement:
Press the MEAS key.
Press the TRANSM CAL softkey.
The R&S FSH now prompts you to connect its RF
input to the tracking generator output so that
calibration can be carried out.
Connect the RF input of the R&S RSH3 directly to
the tracking generator output without the DUT.
Press the softkey F1 or F2 (CONTINUE) to start
calibration.
To abort calibration, press the fourth or fifth softkey
(CANCEL).
During calibration, the R&S FSH outputs the message
"Calibrating THROUGH, please wait..".
Calibration can be aborted by pressing a CANCEL
softkey.
When calibration is over, the R&S FSH outputs the
message "Transm. Calibrated" for 3 seconds.
When calibration is over, the R&S FSH displays
Transm. in the upper right-hand corner of the
measurement diagram. This tells the user that the
R&S
FSH has been calibrated for transmission
measurements. The softkey label TRANSM CAL is
highlighted in green.