Using the R&S FSH in receiver mode
R&S FSH
1145.5973.12 4.64
E-14
Frequency tuning in channel spacings:
For measurement in channel spacings, it is possible to enter a frequency channel instead of the
frequency. A simple channel table can be defined directly from the R&S FSH's front panel. Complicated
channel tables, e.g. ones with gaps in the channel numbers or the frequency, must be defined using the
R&S FSH View software and loaded into the R&S FSH's memory.
Press the FREQ key.
Press the CHANNEL MODE softkey.
The R&S FSH now displays, instead of the frequency, a channel according to the channel table that
was just switched on.
A channel table is selected as follows:
Press the CHANNEL TABLE softkey.
The R&S FSH switches to the submenu for selecting
the channel configuration. All channel tables available
in the instrument that were loaded with the R&S FSH
View software are displayed. If no channel table is
loaded, "No bands available" is displayed.
Select the desired channel table using the rotary
knob or the cursor keys.
Confirm your selection with the SELECT softkey.
The R&S FSH switches to the frequency entry menu.
The frequency is displayed as channels, and the
FREQ softkey is replaced by CHANNEL. All
frequencies are now entered as channel numbers. The
R&S FSH only accepts entries that are defined in the
channel list. Other frequencies can no longer be
entered.
The frequency associated with the set channel is
additionally displayed above the displayed channel.
Set a new channel using the rotary knob or the
cursor keys or enter a new channel via the numeric
keypad.
If you try to enter a channel outside of the defined
range, the R&S FSH displays "Range exceeded".
If no channel table is stored in the instrument or a different one is necessary, a user table can also be
defined.
深圳市盛腾仪器仪表有限公司
Tel:0755-83589391
Fax:0755-83539691
Website: www.Sengt.com