Setting the Amplitude Parameters
1145.5973.12 4.9
E-14
Entering the display range
Press the AMPT key.
Press the RANGE softkey.
A submenu opens. The various options for scaling the
level axis are displayed.
Using the rotary knob or the cursor keys, select the
scaling option you want and confirm by pressing
the ENTER key.
The scaling option you have chosen is immediately set
on the R&S FSH.
The menu items VSWR REFL COEFF (ROH) and
REFL COEFF (mROH) are valid scaling options only if
the R&S FSH is fitted with a tracking generator which
is configured for reflection measurements. The SMITH
CHART menu option is available only if option
R&S FSH-K2 has been installed and a reflection
measurement has been vector-calibrated.
If the R&S FSH-K2 option is installed, the ranges
VSWR 1-1.5 and VSWR 1-1.1 are available in addition.
Entering the display unit
Press the AMPT key.
Press the UNIT softkey.
A submenu opens. The various unit options for the
reference level are displayed.
Using the rotary knob or the cursor keys, select the
unit you want and confirm by pressing the ENTER
key.
The reference level unit is immediately set on the
R&S FSH.