R&S FSH
Trace Settings
1145.5973.12 4.23
E-14
Note:
The memory trace is bit-mapped into the picture memory. Therefore, when the memory
trace is recalled, it will not be adapted to any modifications of the reference level or span
that may have been made in the meantime.
When a stored data set is called, the R&S FSH stores the associated trace in the trace
memory. The stored trace can be displayed with SHOW MEMORY.
Trace mathematics
The R&S FSH can subtract a saved trace from the active trace and display the difference.
Operation:
Press the TRACE key.
Press the TRACE -> MEMORY softkey.
The R&S FSH transfers the currently displayed trace to
the trace memory.
Press the SHOW MEMORY softkey.
The R&S FSH displays the saved trace in white. The
SHOW MEMORY softkey label is highlighted in green
to indicate that the trace in the trace memory is being
displayed.
To remove the saved trace from the screen, press
the SHOW MEMORY softkey again.
Press the TRACE MATH key and select
TRACE – MEM or MEM – TRACE.
The R&S FSH displays the difference between the
saved trace and the active trace.
To remove the saved trace from the screen, press
the TRACE MATH softkey again and select OFF.