Using the R&S FSH in receiver mode
R&S FSH
1145.5973.12 4.66
E-14
Setting the reference level in the amplitude menu:
Press the AMPT key.
Press the REF LEVEL softkey.
Change the reference level setting with the cursor keys or the rotary knob, or enter a new reference
level via the numeric keypad.
Confirm the entry of the reference level by pressing the ENTER key.
Setting the bandwidth
The same bandwidths available in the analyzer mode are also available in the receiver mode. In
addition, the R&S FSH provides the bandwidths 200 Hz, 9 kHz, 120 kHz and 1 MHz for measuring
electromagnetic interference according to CISPR16. In contrast to the analyzer bandwidths, which are
defined as 3 dB bandwidths, the CISPR bandwidths are 6 dB bandwidths.
Press the BW key.
The R&S FSH immediately activates entry of the resolution bandwidth (the MANUAL RES BW softkey
is highlighted in red).
Change the current bandwidth with the cursor keys or the rotary knob, or enter a new bandwidth via
the numeric keypad and terminate the entry with the required unit key.
Note: The 200 kHz bandwidth must always be entered via the numeric keypad.
CISPR16 bandwidths must be set as follows:
Press the MANUAL CISPR BW softkey in the BW
menu.
Select one of the CISPR bandwidths using the
cursor keys or the rotary knob.
Confirm with the ENTER key.
The bandwidths 200 Hz, 9 kHz, 120 kHz and 1 MHz are available.
The CISPR bandwidths are predefined for specific frequency ranges as follows according to CISPR16:
Frequency range
Bandwidth
<150 kHz
200 Hz
150 kHz to 30 MHz
9 kHz
30 MHz to 1000 MHz
120 kHz
>1000 MHz
1 MHz
For this purpose, the R&S FSH automatically sets the predefined bandwidth as a function of the
selected frequency.
Press the BW key.
Press the AUTO CISPR BW softkey.