R&S FSH
Measurements using the power sensor
1145.5973.12 4.71
E-14
Operating sequence:
Press the MEAS key.
Press the MEASURE softkey.
The measurement function submenu opens.
Using the cursor keys or the rotary knob, select the
POWER SENSOR menu item and confirm your
selection with the ENTER key or the MEASURE
softkey.
The R&S FSH opens the screen for power measurements. If no power sensor is connected, no
measured value is displayed. If a power sensor is connected, the R&S FSH sets up a connection via its
interface and after a few seconds displays the measured power.
In the event of incorrect operation or sensor malfunction, the R&S FSH outputs the following error
messages:
Message Cause Remedy
Error in zeroing: signal at
sensor
A signal was present at the
power sensor when zeroing was
performed.
Unscrew the power sensor from
the device under test and repeat
zeroing.
Warning: Input overloaded
The power at the input of the
power sensor exceeds the
permitted power
(23 dBm = 200 mW).
Reduce the power at the sensor
input.
Power sensor hardware error
Communication error between
the R&S FSH and the power
sensor.
Unscrew the sensor from the
R&S FSH and check the
connectors. If the problem
persists, contact a Rohde &
Schwarz service center.
Power sensor error
The power sensor signals an
error to the R&S FSH.
Contact a Rohde & Schwarz
service center.
Unknown power sensor model
connected
The R&S FSH cannot identify the
device connected to its POWER
SENSOR connector.