Using the R&S FSH in receiver mode
R&S FSH
1145.5973.12 4.68
E-14
Measurement on multiple frequencies or channels (scan)
In a scan, the R&S FSH sequentially measures the levels in predefined channels and graphically
displays the measurement results. The length of time the scan dwells on a frequency is determined by
the measurement time. The measurement channels are specified by the selected channel table.
Press the SPAN key.
Press the FREQ SCAN softkey.
The FREQ SCAN softkey is highlighted in green to indicate that the R&S FSH is in the scan mode.
Confirm the entry with the ENTER key.
Press the SCAN START softkey.
Enter the start frequency of the scan via the numeric keypad, or change the start frequency using the
rotary knob or the cursor keys.
Press the STOP SCAN softkey.
Enter the stop frequency of the scan via the numeric keypad, or change the start frequency using the
rotary knob or the cursor keys.
Press the SCAN STEP softkey.
Enter the step size of the scan via the numeric keypad, or change the start frequency using the rotary
knob or the cursor keys.
The R&S now measures on the frequencies defined by
the scan parameters. The levels are displayed by
vertical lines on each of the frequencies. The height of
each line indicates the level.
Pressing the FIXED FREQ softkey switches the R&S FSH back to the default setting of the receiver
mode.
A scan is also possible via the channel table. If frequency entry is set to Channel, the R&S FSH uses
the associated channel table.