R&S ESU
Instrument Functions
Basic Settings
4.246
Operating Manual 1302.6163.12 - 03
• The frequencies/times for each data point must be entered in ascending order,
however, for any single frequency/time, two data points may be input (vertical
segment of a limit line).
The data points are allocated in order of ascending frequency/time. Gaps are not
allowed. If gaps are desired, two separate limit lines must be defined and then
both enabled.
• The entered frequencies/times need not necessarily be selectable in R&S ESU. A
limit line may also exceed the specified frequency or time domains. The minimum
frequency for a data point is -200 GHz, the maximum frequency is 200 GHz. For
the time domain representation, negative times may also be entered. The valid
range is -1000 s to +1000 s.
• The minimum/maximum value for a limit line is -200 dB to +200 dB for the
logarithmic or 10
-20
to 10
+20
or -99.9% to + 999.9% for the linear amplitude scales.
EDIT LIMIT
LINE /
NEW LIMIT
LINE
The
EDIT LIMIT LINE
and
NEW LIMIT LINE
softkeys both call the
EDIT LIMIT LINE
submenu used for editing limit lines. In the table heading, the characteristics of the
limit line can be entered. The data points for frequency/time and level values are
entered in the columns.
Name
Enter
name.
Domain
Select
domain.
Unit
Select
units.
X-Axis
Select
interpolation
Limit
Select upper and lower limit value.
X-Scaling
Entry of absolute or relative values for the x-axis
Y-Scaling
Entry of absolute or relative values for the y-axis
Margin
Entry
of
margin.
Threshold
Entry of vertical threshold (only with relative vertical
scaling)
Comment
Enter
comments.
Time/Frequency
Enter time/frequency for the data points.
Limit/dBm
Enter magnitudes for the data points.
Domain, unit, X scaling and Y scaling cannot be modified as soon as reference
values have been entered in the data section of the table.
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...