Operating Manual 1302.6163.12 - 03
6.181
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]FREQuency:SPAN
0 to f
max
In analyzer mode, this command defines the frequency span.
Example:
"FREQ:SPAN 10MHz"
Characteristics:
*RST value: f
max
with f
max
= maximum frequency
SCPI: conform
Mode:
A
[SENSe<1|2>:]FREQuency:SPAN:FULL
In analyzer mode, this command sets the frequency span to its maximum.
Example:
"FREQ:SPAN:FULL"
Characteristics:
*RST value: -
SCPI: conform
Mode:
A
This command is an event and therefore has no *RST value and no query.
[SENSe<1|2>:]FREQuency:STARt
0 to f
max
This command defines the start frequency of the overall scan in receiver mode. It defines the start
frequency of the sweep in analyzer mode..
Example:
"FREQ:STAR 20MHz"
Characteristics:
*RST value: 0
SCPI: conform
Mode:
R, A-F
[SENSe<1|2>:]FREQuency:STOP
0 to f
max
This command defines the stop frequency of the overall scan in receiver mode. It defines the stop
frequency of the sweep in analyzer mode..
Example:
"FREQ:STOP 20MHz"
Characteristics:
*RST value: f
max
SCPI: conform
Mode:
R, A-F
[SENSe<1|2>:]FREQuency:MODE
CW | FIXed | SWEep | SCAN | TDOMain
In analyzer mode, this command switches between frequency domain (SWEep) and time domain (CW
| FIXed).
In receiver mode, this command switches between frequency scan (SCAN) and time domain scan
(TDOMain).
For CW and FIXed, the frequency setting is made using the
FREQuency:CENTer
command. For
SWEep, the setting is made using the
FREQuency:STARt,
STOP
,
CENTer
and
SPAN
commands.
For SCAN, the frequency setting is made using the
[SENSe<1|2>:]SCAN<1 to 10>:STARt
and
[SENSe<1|2>:]SCAN<1 to 10>:STOP
commands.
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...