Operating Manual 1302.6163.12 - 03
6.51
R&S ESU
Remote Control – Description of Commands
CALCulate Subsystem
6.5.4.1
CALCulate:MARKer:FUNCtion Subsystem
The measurement window is selected by CALCulate 1 (screen A) or 2 (screen B).
CALCulate<1|2>:MARKer<1...4>:FUNCtion:FPEaks[:IMMediate]
<numeric_value>
This command searches the selected trace for the indicated number of maxima. The results are
entered in a list and can be queried with commands
CALC:MARK:FUNC:FPEaks:X?
and
CALC:
MARK:FUNC:FPEaks:Y?
. The number of maxima found can be queried with
CALC:MARK:FUNC:
FPEaks:COUNt?.
The trace to be examined is selected with
CALC:MARK:TRACe
. The order of the
results in the list can be defined with
CALC:MARK:FUNC:FPEaks:SORT
.
Example:
"INIT:CONT OFF"
'switches to single-sweep mode
"INIT;*WAI"
'starts measurement and synchronizes to end
"CALC:MARK:TRAC 1"
'sets marker 1 in screen A to trace 1
"CALC:MARK:FUNC:FPE:SORT X"
'sets the sort mode to increasing X values
"CALC:MARK:FUNC:FPE 3"
'searches the 3 highest maxima for trace 1
"CALC:MARK:FUNC:COUN?"
'queries the number of maxima found
"CALC:MARK:FUNC:Y?"
'queries the level of maxima found
"CALC:MARK:FUNC:X?"
'queries the frequencies (span <> 0) or time (span = 0) of maxima found.
Characteristics:
*RST value: --
SCPI: device-specific
Mode:
A
CALCulate<1|2>:MARKer<1...4>:FUNCtion:FPEaks:COUNt?
This query reads out the number of maxima found during the search. If no search for maxima has been
performed, 0 is returned.
Example:
"CALC:MARK:FUNC:FPE 3"
'searches the 3 highest maxima for trace 1
"CALC:MARK:FUNC:FPE:COUN?"
'queries the number of maxima found
Characteristics:
*RST value: --
SCPI: device-specific
The number of maxima found depends on the waveform and value set for the
Peak Excursion parameter (
CALC:MARK:PEXC
), however, a maximum number of
50 maxima are determined. Only the signals which exceed their surrounding val-
ues at least by the value indicated by the peak excursion parameter will be recog-
nized as maxima. Therefore, the number of maxima found is not automatically the
same as the number of maxima desired.
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...