R&S ESU
Instrument Functions
Analyzer Mode
4.170
Operating Manual 1302.6163.12 - 03
• Spurious emissions (
“SPURIOUS EMISSIONS” on page 4.216
)
• 3
rd
order intercept (
• Harmonic distortion (
“HARMONIC DISTOR” on page 4.214
The above measurements are carried out alternatively.
4.8.13.1
Power Measurement in Time Domain
With the aid of the power measurement function, the R&S ESU determines the
power of the signal in the time domain (SPAN = 0 Hz) by summing up the power at
the individual pixels and dividing the result by the number of pixels. In this way it is
possible to measure for example the power of TDMA signals during transmission or
during the muting phase. Both the mean power and the rms power can be measured
by means of the individual power values.
The result is displayed in the marker info field.
The measured values are updated after each sweep or averaged over a user-
defined number of sweeps (
AVERAGE ON/OFF
and
NUMBER OF SWEEPS
) in
order to determine e.g. the mean power over several bursts. For determination of
the peak value (
MAX HOLD ON
) the maximum value from several sweeps is dis-
played.
Example
Marker info field for:
MEAN
selected,
AVERAGE ON
and
MAX HOLD ON
:
If both the on and off phase of a burst signal are displayed, the measurement range
can be limited to the transmission or to the muting phase with the aid of vertical
lines. The ratio between signal and noise power of a TDMA signal for instance can
be measured by using a measurement as a reference value and after that varying
the measurement range.
Upon switching on power measurement the sample detector is activated (
TRACE -
DETECTOR - SAMPLE)
.
TIME DOM
POWER
MEAN HOLD -2.33 dBm
MEAN AV -2.39 dBm
Side menu
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...