Operating Manual 1302.6163.12 - 03
6.173
R&S ESU
Remote Control – Description of Commands
SENSe Subsystem
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:INSert
AFTer | BEFore
This command inserts a new range before or after the specified range. The range numbers are
updated accordingly. The numeric suffix at RANGe<1...20> specify the range.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Example:
"ESP:RANG3:INS BEF"
'Inserts a new range before Range 3
"ESP:RANG1:INS AFT"
'Inserts a new range after Range 1
Characteristics:
*RST value:-
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:ABSolute:STARt
<numeric_value>
This command defines the absolute limit at the start frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
-400 to +400 dBm
Example:
"ESP:RANG5:LIM:ABS:STAR -40"
'Absolute limit at start frequency in range 5 is -40 dBm
Characteristics:
*RST value: -13dBm
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:ABSolute:STOP
<numeric_value>
This command defines the absolute limit at the stop frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
-400 to +400 dBm
Example:
"ESP:RANG5:LIM:ABS:STOP -40"
'Absolute limit at stop frequency in range 5 is -40 dBm
Characteristics:
*RST value: -13dBm
SCPI: device-specific
Mode:
A
[SENSe<1|2>:]ESPectrum:RANGe<1..20>:LIMit:RELative:STARt
<numeric_value>
This command defines the relative limit at the start frequency of the range for the Spectrum Emission
Mask measurement.
The numeric suffixes <1...20> specify the range. The numeric suffix at SENSe<1|2> is irrelevant.
Parameter:
-400 to +400 dBc
Example:
"ESP:RANG5:LIM:REL:STAR -40"
'Relative limit at start frequency in range 5 is -40 dBc
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...