R&S ESU
Maintenance and Instrument Interfaces
Instrument Interfaces
8.10
Operating Manual 1302.6163.12 - 03
Hardware handshake
For hardware handshake, the instrument indicates that it is ready to receive data via
the lines DTR and RTS. A logic '0' on both lines means 'ready' and a logic '1' means
'not ready'. The RTS line is always active (logic '0') as long as the serial interface is
switched on. The DTR line thus controls the readiness of the instrument to receive
data.
The readiness of the remote station to receive data is reported to the instrument via
the CTS and DSR line. A logic '0' on both lines activates the data output and a logic
'1' on both lines stops the data output of the instrument. The data output takes place
via the TxD line.
Cable for hardware handshake
The connection of the instrument to a controller is made with a so-called zero
modem cable. Here, the data, control and acknowledge lines must be crossed. The
following wiring diagram applies to a controller with a 9-pin or 25-pin configuration.
Fig. 8.54 Wiring of the data, control and acknowledge lines for hardware handshake
Controller
25-pin
Instrument
9-pin
1
2
3
4
5
6
7
8
9
Controlller
9-pin
Instrument
9-pin
1
2
3
4
5
6
7
8
9
1
3
2
6
5
4
8
7
9
8
2
3
6
7
20
5
4
22
RxD / TxD
TxD / RxD
DTR / DSR
GND / GND
DSR / DTR
RTS / CTS
CTS / RTS
RxD / TxD
TxD / RxD
DTR / DSR
GND / GND
DSR / DTR
RTS / CTS
CTS / RTS
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...