R&S ESU
Instrument Functions
Basic Settings
4.302
Operating Manual 1302.6163.12 - 03
The
Save
dialog box offers the following selectable data subsets in the
Items
field.
SELECT ITEMS
The
SELECT ITEMS
softkey sets the focus on the Items field to select the appropri-
ate item to be stored.
ENABLE ALL
ITEMS
The
ENABLE ALL ITEMS
softkey marks all partial data sets.
Remote command:
MMEM:SEL:ALL
DISABLE ALL
ITEMS
The
DISABLE ALL ITEMS
softkey deselects all partial data sets.
Remote command:
MMEM:SEL:NONE
Current Settings
These settings include:
• current configuration of general instrument parameters
• current measurement hardware settings
• active limit lines:
A data set may contain maximum 8 limit lines for each
window. It always contain the activated limit lines and the
de-activated limit lines used last, if any. Consequently, the
combination of the restored deactivated limit lines depends
on the sequence of use with command
MMEM:LOAD
.
• the activated transducer factor
• user-defined color settings
• configuration for hardcopy output
• active transducer set:
A data set may contain maximum 4 transducer factors. It
always contain the activated factors and the factors used
and de-activated last, if any. Consequently, the combination
of the restored deactivated transducer factors depends on
the sequence of use with the command
MMEM:LOAD
.
• peak list or final measurement results
(receiver mode)
• settings of tracking generator
(only with option R&S ESU B9)
All Limit Lines
all
limit
lines
All Transducer
all transducer factors
All Transducers
all
transducer
All Traces
all traces which are not blanked
Source Cal Data
correction data for tracking generator
(only with options R&S ESU B9 or R&S ESU B10)
Scan Channel List
adds the channel lists to the list of data subsets of a save/
recall device setting.
(only with option R&S ESU K50)
Peak-Final List
peak list or final measurement results
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...