![R&S ESU Series Скачать руководство пользователя страница 540](http://html1.mh-extra.com/html/rands/esu-series/esu-series_operating-manual_3138875540.webp)
Operating Manual 1302.6163.12 - 03
6.101
R&S ESU
Remote Control – Description of Commands
DIAGnostic Subsystem
DIAGnostic<1|2>:SERVice:INPut:RECTangle[:STATe]
ON | OFF
This command switches the rectangle calibration signal on or off in the selected measurement window
(suffix <1|2>). The setting takes effect only if the RF input is set to the internal reference signal (
DIAG:
SERV:INP CAL
command).
Example:
"DIAG:SERV:INP CAL"
"DIAG:SERV:INP:RECT ON"
Characteristics:
*RST value: OFF
SCPI: device-specific
Mode:
R, A
DIAGnostic<1|2>:SERVice:INPut:RECTangle:PRATe
5 kHz | 31.25 kHz | 50 kHz | 250 kHz | 500 kHz
This command sets the frequency of the rectangle calibration signal in the selected measurement
window (suffix <1|2>).
5 kHz | 31.25 kHz | 50 kHz | 250 kHz | 500 kHz can be set.
Example:
"DIAG:SERV:INP:RECT:PRAT 128 MHz
Characteristics:
*RST value: 5 kHz
SCPI: device-specific
Mode:
R, A
DIAGnostic<1|2>:SERVice:SFUNction
’<string>’
This command activates a service function which can be selected by indicating the five parameters:
function group number, board number, function number, parameter 1 and parameter 2 (see service
manual). The contents of the parameter string is identical to the code to be entered in the data entry
field of manual operation.
The entry of a service function is accepted only if the system password Level 1 or Level 2 has been
entered previously (command:
SYSTem:SECurity
).
The numeric suffix <1|2> is ignored with this command.
Example:
"DIAG:SERV:SFUN '2.0.2.12.1'"
Characteristics:
*RST value: -
SCPI: device-specific
Mode:
R, A
If the rectangle calibration signal is switched on, the pulsed calibration signal is
switched off.
The service functions of the instrument are not identical to those of the FSE family.
That is why the remote command differs in syntax and data format.
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...