R&S ESU
Instrument Functions
Analyzer Mode
Fig. 4.13 Display of the complementary cumulative distribution function (CCDF)
Alternate to the histogram display of the APD the Complementary Cumulative Distri-
bution Function (CCDF) can be displayed. It shows the probability of an amplitude
exceeding a specific value.
For the APD function the x-axis is scaled in absolute values in dBm, whereas for the
CCDF function the x-axis is scaled relative to the MEAN POWER measured.
Definitions:
Crest factor = peak voltage to rms
CCDF = complementary cumulative distribution function
During an active statistic measurement, the functions FULL SCREEN,
SPLIT SCREEN and selection of the active diagram via the SCREEN A / SCREEN
B hotkeys are disabled.
Содержание ESU Series
Страница 1: ...R S ESU EMI Test Receiver Operating Manual 1302 6163 12 03 Test Measurement Operating Manual...
Страница 25: ......
Страница 27: ...R S ESU Putting into Operation 1 2 Operating Manual 1302 6163 12 03...
Страница 29: ...R S ESU Getting Started 2 2 Operating Manual 1302 6163 12 03...
Страница 31: ...R S ESU Manual Operation 3 2 Operating Manual 1302 6163 12 03...
Страница 349: ...R S ESU Instrument Functions Basic Settings 4 318 Operating Manual 1302 6163 12 03...
Страница 405: ...R S ESU Instrument Functions User Port 4 374 Operating Manual 1302 6163 12 03...
Страница 439: ...R S ESU Remote Control Basics Status Reporting System 5 34 Operating Manual 1302 6163 12 03...
Страница 821: ...R S ESU Remote Control Programming Examples More Complex Programming Examples 7 32 Operating Manual 1302 6163 12 03...
Страница 841: ...R S ESU Error Messages Device Specific Messages 9 10 Operating Manual 1302 6163 12 03...
Страница 854: ...Operating Manual 1302 6163 12 03 10 13 R S ESU...
Страница 855: ...R S ESU 10 14 Operating Manual 1302 6163 12 03...