- 203 -
Frame format:
1 2-3
4
5
6
7-15
16 17 18 19 20 21
S space stability
marker space character mass space
unit
CR LF
Example:
S CR LF
– command sent from a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
- command carried out, response
is mass value in basic measuring unit.
Immediately send measurement result in basic measuring unit
Format:
SI CR LF
Accessible responses:
SI_I CR LF
- command understood but not accessible at this
moment
MASS FRAME
- response is immediate with mass value in basic
weighing unit
Frame format:
1 2
3
4
5
6
7-15
16 17 18 19 20 21
S I space stability
marker space character mass space
unit
CR LF
Example:
S I CR LF
– command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF
- command carried out,
immediate response of mass value in basic weighing unit
Immediately send measurement results from all weighing platforms
basic measuring units
Format:
SIA CR LF
Accessible responses:
SIA_I CR LF
- command understood but not accessible at
this moment
MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF
- immediate response of mass value in basic
measuring unit from all weighing platforms
Frame format with mass from each weighing platform:
1 2
3
4
5
6
7-15
16 17 18 19 20 21
P n space stability
marker space character mass space
unit
CR LF