- 89 -
product record of the database of products. To use this function,
before starting the density determining process, using database of
product select a product for which the density procedure is carried
out (if density has already been determined for this product, it is
replaced with new value),
o
Tare mode,
o
Printout mode / Value release,
o
Printout
Means of using the above functions are given in point 10.8. “Additional
parameters on weighing process”.
20.7.
Printouts
Option Printouts enables setting the contents of printouts in a standard
printout and also determining non-standard printouts.
Standard printout
comprises four internal blocks, each featuring different variables. Each
variable can be attributed with parameter YES – i.e. present on a
printout, or NO – i.e. not printed.
Settings for header printout, weighing printout, and footer printout are specified in
point. 15.5; below description refers to setting a printout template from density mode.
A user can design content of a report for each of carried out procedures. On
pressing <Density printout template> option a window is opened enabling setting
the content of each report.
Content of each of reports:
o
SOLID
−
Working mode
−
Procedure
−
Sample no.
−
User
−
Balance type
−
Balance ID