- 195 -
ambient conditions at a workstation. The function provides a user with
data on the best possible balance settings for obtaining the highest
measurement repeatability and the shortest measurement time with
acceptable value of repeatability.
Test results are not stored in balance’s memory.
Exiting function causes erasing the test results.
Additionally the function enables:
•
Printing test results on printers available in the system and selecting
the most optimum parameters from the balance’s option level.
Measurement results:
*Filter type
*Setting of “Value release” parameter
*Repeatability of indications expressed as standard deviation
*Average time of measurement stabilization
Example of a report:
-------- Autotest Filter: Report --------
Balance type PS 3Y
Balance ID 400010
User Admin
Software rev. L0.0.21 S
Date 2012.01.16
Time 10:17:27
----------------------------------------
Reading unit 0.001/0.01 g
Internal weight mass 1402.094 g
----------------------------------------
Filter Very fast
Value release Fast
Repeatability 0.0042 g
Stabilization time 4.505 s
Filter Very fast
Value release Fast & Reliable
Repeatability 0.0097 g
Stabilization time 4.524 s
Filter Very fast
Value release Reliable
Repeatability 0.0092 g
Stabilization time 4.895 s
Filter Fast
Value release Fast
Repeatability 0.0099 g
Stabilization time 4.979 s
.
.