- 88 -
9.
Load aluminium mass standard on balance’s weighing pan and on
stabilization of measurement result accept it by pressing <
> key,
10.
The balance automatically calculates the density of air, and indicates its
value on the display,
11.
Press <
> key to finish the procedure.
The value of determined air density is automatically assigned to field <Air
density> in menu <Settings/ Air buoyancy compensation> in working mode:
weighing.
20.6.
Additional settings of Density mode
The additional settings enable adjusting the working mode to user’s needs and
requirements. To access the setting follow below procedure:
Procedure:
1.
Press grey coloured workspace,
2.
The display indicates menu of: Settings, Keys, Information, Printout
and Profile,
3.
Press <Settings> key,
4.
The display indicates functions related to density determining
process, such as:
o
Request sample number
Available settings:
NO
– data on sample number is not required; applicable in case of
a single measurement
YES
– before each measurement a window requesting inserting
sample number is automatically displayed; applicable in case of a
series of measurements
o
Measuring unit: the balance enables choosing between the
following units: [g/cm
3
], [kg/m
3
], [g/l]. Selected unit is activated
for all density determining options and also in final printout.
o
Assign density to a product: YES/NO – if option <YES> is
selected, the software automatically assigns the value of
determined density of solid or liquid to field <Density> in a