![RADWAG 3Y Series Скачать руководство пользователя страница 207](http://html1.mh-extra.com/html/radwag/3y-series/3y-series_user-manual_3937928207.webp)
- 207 -
Set mass value of a single (only for PARTS COUNTING)
Format:
SM_XXXXX CR LF
, where:
_ -
space,
XXXXX
– mass format
Accessible responses:
SM_OK CR LF
SM_I CR LF
- command carried out
- command understood but not accessible at this
moment (e.g.: other working mode, not PARTS
COUNTING)
ES CR LF
- command not recognized (incorrect mass format)
Set target mass value (e.g. for DOSING)
Format:
TV_XXXXX CR LF
, where:
_ -
space,
XXXXX
– mass format
Accessible responses:
TV_OK CR LF
TV_I CR LF
- command carried out
- command understood but not accessible at this
moment (e.g.: WEIGHING mode)
ES CR LF
- command not recognized (incorrect mass format)
Set reference mass value (e.g. for PERCENT SETUP)
Format:
RM_XXXXX CR LF
, where:
_ -
space,
XXXXX
– mass format
Accessible responses:
RM_OK CR LF
RM_I CR LF
- command carried out
- command understood but not accessible at this
moment (e.g.: other working mode, not PERCENT
SETUP)
ES CR LF
- command not recognized (incorrect mass format)
User log in
Format:
LOGIN Name, Password CR LF
(Name and password have to be
inserted as they are set in the balance – small and capital letters)
Accessible responses:
LOGIN OK CR LF
- command understood, new user is logged in
LOGIN ERRROR CR LF - command understood, an error in name or
password occurred, log in failed
ES CR LF
- command not recognized (error in format)