- 141 -
•
List of data for internal criteria
Internal control
Internal control criteria on <YES> / off <NO>
Sample quantity
Sample quantity value for a product
Error value [- T]
Error value of marginal negative
T
(-T) given
in unit of product nominal mass
Qn
.
Measurements lower than value of
Qn-T
are
regarded as valid.
Error value [+ T]
Error value of marginal positive
T
(+T) given
in unit of product nominal mass
Qn
.
Measurements higher than value of
Qn+T
are
regarded as valid.
Quantity of smaple
[Qn-2T] dis.
Number of negative errors (
–2T)
of a tested
sample which disqualifies control
Quantity of smaple
[Qn-2T] dis.
Number of positive errors (
+2T)
of a tested
sample which disqualifies control
Quantity of smaple
[Qn-T] dis.
Number of negative errors (
–T)
of a tested
sample which disqualifies control
Quantity of smaple
[Qn+T] dis.
Number of positive errors (
+T)
of a tested
sample which disqualifies control
Average limit value
Mode for estimation of average limit value
(constant or automatic)
Average limit value -
Value of average limit (negative) of a tested
sample (concerns value of average limit as
“constant”)
Average limit value +
Value of average limit (positive) of a tested
sample (concerns value of average limit as
“constant”)
Factor value [-Wk]
Multiplier of standard deviation for value of
average
limit
(negative)
calculated
automatically
Factor value [+Wk]
Multiplier of standard deviation for value of
average
limit
(positive)
calculated
automatically