- 128 -
25.5.
Copy tare
The option enables copying once selected tare value in a series to all samples
which do not have tare value assigned, and for which differential weighing
process (status other than Weighing B) is not completed.
Means of operation:
Press key
Which displays list of samples for which
tare value is assigned.
Press tare value that should apply to
other samples in a series (which do not
have tare applied).
The software automatically applies
selected tare value to data of these
samples.
25.6.
Using option “SAMPLE SELECTION”
The option enables selecting a sample that is prepared for weighing, with no
need to weigh samples is a sequence as suggested by the software. The option is
available only after initiating a given process.
Means of operation:
While in differential weighing mode, on selecting a series, press procedure
selecting key, e.g. <Weighing A>.
Then press the following key:
which opens a list of samples for which
the selected procedure can be carried
out.
Select a sample to be weighed from
displayed list.
This option is particularly useful if a
series contains a large number of
samples, and the user should weigh a
sample that is at the end of a series.